A3 AAS Atomic Absorbtions Spectrometer

The A3 Atomic Absorption Spectrometer is a high performance automated instrument designed to meet the requirements of the modern laboratory. Due to its versatility and performance it can be used for a wide range of applications.
Three flame options are available to the user with the Air/Acetylene being the standard configuration. This flame can be used for nearly all standard elements while the N2O/Acetylene and the Air/LPG (natural gas) are available as an option for the more demanding of elements. All three flame configurations offer coded burner for full safety protection.
The graphite head is fixed into the optical path to maximize performance and eliminate drift. The temperature of the transversely heated graphite tube is accurately controlled by means of a precision feedback system. Pyrolytically coated platform tubes are supplied as standard to improve the performance and eliminate many analytical problems associated with this technique.
The flame and graphite furnace can be changed over by a simple selection in versatile AA Win Pro software.


Key Features
  • Easy to use, complete software control
  • Integrated safety precautions for user protection
  • Flexible configuration for a wide array of industrial apps
  • Time-saving high speed analysis

Technical Specifications

Model:Flame AA System-A3F, Graphite AA System-A3G, 

Combined Flame and Graphite Furnace AA System-A3FG

Spectrometer System
Wavelength range190-900 nm
Light sourceHollow cathode lamp (HCL), Deuterium Arc lamp (D2)
Modulation Square wave pulse
Modulation frequency

100 hz self reversal (SR) background,

 400 hz Deuterium Arc (D2) background

Monochromator 
Grating 1800 grooves/mm diffraction grating
Blazed wavelength250 nm
Focus 300 nm
Bandwidth 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm
Scan modeAutomatic
Photometric typeSingle beam
Wavelength accuracy±0.15 nm
Wavelength resolution0.2 nm±0.02 nm
Wavelength reproducibility<0.05 nm
Spectral bandwidth0.5, 1.0, 2.0, 5.0 nm
Baseline stability0.005 A/30 min
Stray light≤0.12 %T (220 nm NaI, 340 nm NaNO2)
Background correction

Deuterium Arc (D2) 1.0 Abs,

Self Reversal (SR) 3.0 Abs

Flame Analysis
Flame types

Air/Acetylene,

Nitrous Oxide/Acetylene,

Air/Propane (LPG)

Sensitivity (Cu)2 μg/ml Absorption >0.280 Abs
Characteristic concentration

Cu<0.02 μg/ml,

Ba<0.15 μg/ml (N2O/Acetylene)

Detection limitCu<0.004 μg/ml
RepeatabilityCu<0.7 % (Air/Acetylene Flame)
    Ba<1.0 % (Nitrous Oxide/Acetylene Flame)
Burner headsTitanium alloy
Nebulizer 

High efficiency glass,

Acid proof available as an option

Pre-mix chamberCorrosion resistant
Atomizer selectionManual (A3F)
    Automatic changeover (A3FG)
Safety featuresBurner identification, flame sensor, gas leak sensor, low gas pressure sensor, drain trap sensor, power loss protection
Graphite Furnace Analysis 
Graphite head Transversally heated
Temperature rangeAmbient-2650 ℃
Heating programUp to 10 steps. Drying, Ashing, Atomization, Cleaning 
Feedback Voltage and optical temperature control feedback
Sensitivity (Cu)50 ng/m Absorption>0.40 Abs
Detection limit Cd<0.004 ng/ml
Repeatability

Cu<2.0 %,

Cd<2.0 %

Graphite tubesPyrolytically coated with platform
Sample sizeUp to 20μl
Graphite coolingWater circulator available
Safety features

Burner Identification,

Flame Sensor,

Gas leak Sensor,

Low Gas Pressure Sensor,

Drain Trap Sensor,

Power Loss Protection,

Circulation Water (graphite),

Over Temperature Sensor (graphite)

Data Processing 
Operating programAA Win Pro software
Analytical methods

Flame AA, Flame AE,

Graphite Furnace,

Hydride Generation

Readout ModeContinuous, manual, peak height, peak area
calibrationsMulti-standard calibration, standard addition, interpolation
Data storageAnalytical results, instrument and measurement parameters, signal profile, calibration curve
Power Requirements
Main unit 220 VAC 50/60 Hz 200 W
Graphite power supply

380VAC 50/60Hz,

instantaneous power 5 kW current rating 7 kW

Dimensions 
Main unit

44 x 21 x 21 in

165 lb

Graphite power supply

20 x 21 x 21 in

154 lb


A3 AAS Atomic Absorbtions Spectrometer-Brochure

Accessories & Spare Parts

  • AS3 Universal Auto sampler


    Description:

    The device is a dedicated flame and graphite auto sampler for A3 Atomic Absorption Spectrophotometer. It can suck standard samples and samples automatically and clean the sampling head and pipelines automatically. Also can prepare standard by auto dilution.


  • Single Element Hollow Cathode Lamps (All Elements)


    Description:

    Different element analysis needs different element lamp.


  • 0932-04-03-01-00 Glass Nebulizer


    Description: 

    For general purpose, cannot proof acid.

  • B301001-00 Graphite Tubes with Platform


    Description: 

    Life time over than 300 times heating

  • CW-1Y Cooling Water Circulator


    Description:

    CW-1Y Cooling water circulator takes water as

    the heat transfer medium to transfer the heat

    quantity produced by the instruments or equipments,

    then the refrigerating machine dissipates

    the heat outside to ensure the instruments

    working in the normal temperature.


  • AC-1Y Air Compressor


    Description:

    AC-1Y air compressor can be used to supply an atomic absorption spectrophotometer with compressed air.

    Rated discharge pressure: 44psi

    Output rate: 0.3-0.9m3/h (adjustable)

    Noise: smaller than 55dB

    Power supply: AC110V 60HZ

    Dimension: 400×300×620(mm)


  • Flame Burner head


    Description:

    500-023 Coded Nitrous Oxide Burner 50mmTi Alloy Burner

    500-030 Coded Propane Burner 3*100mm Ti Alloy Burner


 

Query Form(*All fields are mandatory)